Possible Faults

Fault

Possible cause

Possible Solution

0320_devsetup_empty

In the context menu Add Device the above symbol is displayed next to the TempScan.

The device type has not been activated.

Activate the device type TempScan (See operating manual of the PowerView3).

The TempScan20 cannot be configured.

The baud rate of the TempScan20 does not correspond to that of the PowerView3.

Change the baud rate of the TempScan20 (refer to PowerView3 operating manual).

 

The TempScan20 has been assigned with the wrong node ID in the PowerView3.

Enter the correct node ID for your TempScan20 in the PowerView3 (refer to PowerView3 operating manual).

Measured values are incorrect.

Wrong thermocouple type settings

Enter the correct thermocouple type in the PowerView3 (see section Settings in the PowerView3).

The thermocouple is faulty.

Replace the thermocouple (see section Connecting Thermocouples).

A thermocouple does not display any measured values.

The relevant channel is not activated.

Activate the relevant channels in the PowerView3 (see section Settings in the PowerView3).

The thermocouple is faulty.

Replace the thermocouple (see section Connecting Thermocouples).

The thermocouple is not connected properly.

Connect the thermocouple correctly (see section Connecting Thermocouples).

A threshold value cannot be set to the desired temperature.

Threshold values are too close to each other.

Threshold values must differ by a minimum of 10 °C (18 °F).